RRC ID 66045
著者 Takano W, Shibata S, Hagen N, Matsuda M, Otani Y.
タイトル Minimizing scattering-induced phase errors in differential interference contrast microscopy.
ジャーナル J Biomed Opt
Abstract SIGNIFICANCE:Differential interference contrast (DIC) microscopes allow noninvasive in vivo observation of transparent microstructures in tissue without the use of fluorescent dyes or genetic modification. We show how to modify a DIC microscope to measure the sample phase distribution accurately and in real-time even deep inside sample tissue.
AIM:Our aim is to improve the DIC microscope's phase measurement to remove the phase bias that occurs in the presence of strong scattering.
APPROACH:A quarter-wave plate was added in front of the polarization camera, allowing a modified phase calculation to incorporate all four polarization orientation angles (0 deg, 45 deg, 90 deg, and 135 deg) captured simultaneously by the polarization camera, followed by deconvolution.
RESULTS:We confirm that the proposed method reduces phase measurement error in the presence of scattering and demonstrate the method using in vivo imaging of a beating heart inside a medaka egg and the whole-body blood circulation in a young medaka fish.
CONCLUSIONS:Modifying a polarization-camera DIC microscope with a quarter-wave plate allows users to image deep inside samples without phase bias due to scattering effects.
巻・号 25(12)
公開日 2020-12-1
DOI 10.1117/1.JBO.25.12.123703
PII JBO-200280SSRR
PMID 33319525
PMC PMC7734411
MeSH Animals Microscopy* Microscopy, Interference
IF 2.785
リソース情報
メダカ IB178 HdrR-II1 MT827 T5